Western Society of Periodontics

Laboratory Studies

Volume Number 4, 1995


Back to Index

Evaluation of periodontal scalers

An automated scaling apparatus that simulated the scaling process of hand instrumentation was developed to quantitatively analyze the cutting quality and abrasive resistance of scalers. The effects of lateral scaling force (700, 500, and 300 dyne) and scaler angles (70 degrees to 95 degrees) on the cutting quality and abrasive resistance of scalers were evaluated quantitatively by the amount of the abraded material worn away in 1,000 strokes. Comparison of the three scaling forces showed a greater amount of abrasion at higher force than that at lower force. This suggests that the decrease in the amount due to abrasion could be compensated by increasing the lateral scaling force. Regarding the scaler angle, results indicated that the amount of material removed increased with an increase of the scaler angle up to 70 degrees, but then rapidly decreased at an angle of 90 degrees or more. The most effective scaling angle was 87 degrees, and this was not affected by scaling force. These results suggest that a greater amount of removal could be obtained at a scaling angle of 87 degrees and a scaling force of 700 dyne. The present findings suggested the automatic scaling apparatus could be a useful tool for quantitatively evaluating the cutting quality and abrasive resistance of scalers. [P.W.]

Kaya, H., T. Fujimura, and S. Kimura, J Periodont, 66:62, 1995